1.Portable structure design, easy to use
2.All stainless steel chassis, easy to clean, suitable for use in clean areas
3.Tilted control panel for easy observation and operation
4.Test gas is self-supplied
5.Provide result printing function, easy to save records
6.Use SIEMENS programmable controller (PLC) and touch screen for control
7.Leak test on sleeves and gloves before or after use of the isolator
Input pressure: 0.2MPa~0.6MPa
|
Inflflation pressure: 500~1000Pa
|
Test time: 5-8min
|
Pressure display resolution: 0.1Pa
|
Detection accuracy (minimum aperture): 500μm
|
Instrument weight: 17k
|
Test port specififications: customized (round, oval)
|
Power supply battery: 220V
|
Application area: used to test isolator/RABS;
|
system sleeve/glove or integrated glove integrity
|
Size:303×500×410mm
|